{"product_id":"ametek-mg-dm-t-150-1l-c-deadweight-tester-10-15000-psi","title":"Ametek MG DM-T-150-1L\/C Deadweight Tester, 10-15,000 psi","description":"\u003ch2\u003eAmetek MG DM-T-150-1L\/C Deadweight Tester\u003c\/h2\u003e\u003ch3\u003eOverview\u003c\/h3\u003e\n\u003cp\u003eThe Ametek MG DM-T-150-1L\/C Deadweight Tester is a Type T hydraulic deadweight tester from Ametek's M\u0026amp;G (Mansfield \u0026amp; Green) line, designed to bring laboratory accuracy into industrial environments. This dual-column, dual-piston configuration covers two pressure ranges, 10 to 1,500 psi and 100 to 15,000 psi (to 1,000 bar), with 5 psi and 50 psi weight increments respectively, so range changes need no piston exchange. The -1L\/C suffix indicates certification to local gravity with an accuracy of ±0.025% of the indicated reading.\u003c\/p\u003e\n\u003cp\u003eEach tester features a re-entrant type measuring piston and cylinder assembly that reduces fluid leakage and increases float time, helping maintain accuracy as test pressures rise. Construction uses 300 series stainless steel and Monel with Buna-N O-rings standard (Viton and EPT optional), and the tester can run on a distilled water\/alcohol mix or hydraulic oil as the pressure medium. A lever-actuated, dual-volume hydraulic pump generates pressure, and testers are manufactured to the International Standard Gravity or to a user-specified local gravity.\u003c\/p\u003e\n\u003ch3\u003eFeatures\u003c\/h3\u003e\n\u003cul\u003e\n\u003cli\u003eDual pressure ranges: 10 to 1,500 psi and 100 to 15,000 psi\u003c\/li\u003e\n\u003cli\u003e±0.025% of reading accuracy, certified to local gravity (1L\/C configuration)\u003c\/li\u003e\n\u003cli\u003eDual-column, dual-piston Type T hydraulic deadweight tester\u003c\/li\u003e\n\u003cli\u003eRe-entrant piston\/cylinder design for low leakage and long float time\u003c\/li\u003e\n\u003cli\u003eLever-actuated, hand-operated hydraulic pump with dual volume control\u003c\/li\u003e\n\u003cli\u003e300 series stainless steel and Monel wetted parts\u003c\/li\u003e\n\u003cli\u003eWater\/alcohol mix or AAA hydraulic oil pressure medium\u003c\/li\u003e\n\u003cli\u003eNIST-traceable calibration certificate supplied\u003c\/li\u003e\n\u003c\/ul\u003e\n\u003ch3\u003eApplications\u003c\/h3\u003e\n\u003cul\u003e\n\u003cli\u003ePrimary calibration of pressure gauges, transmitters, and transducers\u003c\/li\u003e\n\u003cli\u003eLaboratory pressure standards and plant calibration benches\u003c\/li\u003e\n\u003cli\u003eVerification of test gauges in manufacturing environments\u003c\/li\u003e\n\u003c\/ul\u003e\n\u003ch3\u003eAmetek MG DM-T-150-1L\/C Deadweight Tester Technical Specifications\u003c\/h3\u003e\n\u003ctable\u003e\u003ctbody\u003e\n\u003ctr\u003e\n\u003ctd\u003eType\u003c\/td\u003e\n\u003ctd\u003eHydraulic piston gauge (Type T deadweight tester)\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003ePressure range\u003c\/td\u003e\n\u003ctd\u003e10 to 1,500 psi and 100 to 15,000 psi (to 1,000 bar)\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003eIncrements\u003c\/td\u003e\n\u003ctd\u003e5 psi (low range); 50 psi (high range)\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003eAccuracy\u003c\/td\u003e\n\u003ctd\u003e±0.025% of indicated reading\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003eRepeatability\u003c\/td\u003e\n\u003ctd\u003e±0.005% of indicated reading\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003eConfiguration\u003c\/td\u003e\n\u003ctd\u003eDual column, dual piston\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003eGravity basis\u003c\/td\u003e\n\u003ctd\u003eLocal gravity (L), with data (C suffix)\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003eUnits\u003c\/td\u003e\n\u003ctd\u003epsi, kg\/cm², bar, kPa available\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003ePump\u003c\/td\u003e\n\u003ctd\u003eLever action, hand-actuated, dual volume hydraulic pump\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003eWetted parts\u003c\/td\u003e\n\u003ctd\u003e300 series stainless steel, Monel; Buna-N O-rings standard\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003eMedium\u003c\/td\u003e\n\u003ctd\u003eDistilled water\/alcohol mix (standard) or AAA hydraulic oil\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003eConnections\u003c\/td\u003e\n\u003ctd\u003e1\/4 in. and 1\/2 in. NPT female pressure connections\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003eWeight material\u003c\/td\u003e\n\u003ctd\u003eHard, non-magnetic die-cast alloy\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003c\/tbody\u003e\u003c\/table\u003e\n\n\u003cp\u003eManufacturer information: \u003ca href=\"https:\/\/www.ametekcalibration.com\/-\/media\/ametekcalibration\/download_links\/pressure\/deadweight%20testers\/deadweight-tester-precision-pressure-measurement-guide-us.pdf\" target=\"_blank\" rel=\"noopener\"\u003eAMETEK Deadweight Testers documentation\u003c\/a\u003e\u003c\/p\u003e","brand":"Ametek","offers":[{"title":"Default Title","offer_id":67776342524091,"sku":"IMM_18893852","price":8680.0,"currency_code":"USD","in_stock":true}],"thumbnail_url":"\/\/cdn.shopify.com\/s\/files\/1\/0777\/2635\/1547\/files\/ametek-mg-dm-t-150-1l-c.png?v=1792552898","url":"https:\/\/na.indomultimeter.net\/products\/ametek-mg-dm-t-150-1l-c-deadweight-tester-10-15000-psi","provider":"Indo Multimeter","version":"1.0","type":"link"}