AccuMac AM1730-9 Secondary Reference PRT
$266.00
AccuMac AM1730-9 Secondary Reference PRT
Overview
The AM1730-9 is the 9-inch (228.6 mm) member of the AM1730 series of secondary reference PRTs. A specially supported platinum sensing element shields the wire from contamination at high temperature, giving short-term stability below ±0.007 °C with low hysteresis and thermal-shock effects.
Each probe fully meets the ITS-90 criteria for reference thermometers and ships with an ISO 17025 accredited calibration. The Inconel sheath withstands harsh environments, and Teflon-insulated four-lead copper wire runs 2 m from the handle.
Features
- Calibrated accuracy ±0.015 °C at 0.01 °C (k=2)
- Short-term stability below ±0.007 °C
- Temperature range -200 to 420 °C
- 100 Ω nominal resistance, 0.003925 Ω/Ω/°C coefficient
- 0.188 in OD × 9 in Inconel sheath
- Thermal shock below ±0.005 °C after 10 thermal cycles
- Hysteresis ≤0.005 °C
- ISO 17025 accredited calibration included
Applications
- Reference standard for dry-well and bath calibration
- Precision temperature measurement in laboratory and field
- Calibration of industrial PRTs and thermometers
AccuMac AM1730-9 Secondary Reference PRT Technical Specifications
| Temperature range | -200 to 420 °C |
|---|---|
| Nominal resistance at 0.01 °C | 100 Ω |
| Temperature coefficient | 0.003925 Ω/Ω/°C |
| Calibrated accuracy (k=2) | ±0.025 °C at -40 °C; ±0.015 °C at 0.01 °C; ±0.035 °C at 420 °C |
| Long-term drift | <0.01 °C at TPW after 100 hours at maximum temperature |
| Short-term stability | <0.007 °C |
| Thermal shock | <0.005 °C after 10 thermal cycles |
| Hysteresis | ≤0.005 °C |
| Self-heating | 50 mW/°C |
| Response time | 6 seconds (63% step in water at 1 m/s) |
| Sensor | 32 mm long, 3 mm from tip |
| Sheath | 0.188 in (4.775 mm) OD × 9 in (228.6 mm) Inconel |
| External leads | 4 leads, Teflon insulated, 2 m |
| Measurement current | 0.5 mA or 1 mA |
Manufacturer information: AM1730 on accumac.com.