Additel ADT875-1210-AR-110V Temperature Calibrator

Overview

The ADT875-1210 is Additel's standard 1210 °C furnace in the 87X-1210 series. Advanced Adaptive Control combines patent-pending wind tunnel technology with 3-zone temperature control to deliver high uniformity and stability, and each unit is tested and calibrated in Additel's ISO 17025 accredited laboratory before shipment, with the accredited certificate and data included.

Metallic interchangeable inserts such as the supplied AR insert let one furnace serve different probe sizes, while Wi-Fi, Bluetooth and a colour touch screen handle operation and setup. An optional on-board process calibrator (PC option) adds thermocouple readout, switch testing and transmitter work with HART communication, and an external control probe input enables a self-calibration routine.

Features

  • Temperature control from 100 to 1210 °C
  • Display accuracy ±1.2 °C at 100 °C, ±2.0 °C at 1210 °C
  • Stability ±0.1 °C
  • Axial uniformity ±0.6 °C at 100 °C, ±1.5 °C at 1210 °C
  • Radial uniformity ±0.2 °C at 100 °C, ±1.0 °C at 1210 °C
  • 3-zone Advanced Adaptive Control
  • Metallic interchangeable insert (AR) in a 24.8 mm well
  • Wi-Fi, Bluetooth and colour touch screen

Applications

  • Thermocouple calibration and verification work
  • Calibrating thermocouples, switches and transmitters with the PC option
  • Laboratory heat source for high-temperature sensor testing

Additel ADT875-1210-AR-110V Temperature Calibrator Technical Specifications

Temperature range 100 to 1210 °C
Display accuracy ±1.2 °C at 100 °C; ±2.0 °C at 1210 °C
Stability ±0.1 °C
Axial uniformity (20 mm zone) ±0.6 °C at 100 °C; ±1.5 °C at 1210 °C
Radial uniformity ±0.2 °C at 100 °C; ±1.0 °C at 1210 °C
Loading effect ±0.5 °C (1-4 probes, excluding sensor)
Heating time 50 min (23 to 1210 °C)
Cooling time 50 min to 300 °C; 50 min to 50 °C
Typical time to stability 15 min
Resolution 0.01 °C
Well diameter 24.8 mm OD
Control 3-zone Advanced Adaptive Control
Connectivity Wi-Fi, Bluetooth, colour touch screen
Power 110 V
Certificate ISO 17025 accredited calibration with data

Manufacturer information: Additel 87X-1210 on additel.com.