Skip to product information
Additel ADT878-160-A-220V Reference Dry Well, -40 to 160 °C, Insert A
$7,532.00
Additel ADT878-160-A-220V Reference Dry Well
Overview
The ADT878 series brings dual-zone control, a self-calibration feature that lets you recalibrate the internal control sensor against an external reference, and set point control by reference. The ADT878-160 also supports an optional TPW kit for automatic realization and maintenance of a triple point of water cell.
A 160 mm immersion depth, 31.9 mm insert OD and 10-minute typical time to stability support day-to-day thermometer calibration. Communication is provided over USB A, USB B, RJ45, Wi-Fi and Bluetooth.
Features
- Temperature range -40 to 160 °C at 23 °C ambient
- Display accuracy ±0.1 °C at full range
- Stability ±0.005 °C (30 min)
- Radial uniformity ±0.01 °C at full range
- Self-calibration of the internal control sensor
- Dual-zone control and set point by reference
- 6.5-inch color touch screen, 0.001 °C resolution
- USB, RJ45, Wi-Fi and Bluetooth communication
Applications
- Laboratory thermometer calibration
- PRT and thermocouple verification
- Triple point of water realization with TPW kit (optional)
Additel ADT878-160-A-220V Reference Dry Well Technical Specifications
| Model | ADT878-160 (non-PC, insert A, 220 V) |
|---|---|
| Temperature range | -40 to 160 °C at 23 °C ambient |
| Display accuracy | ±0.1 °C at full range |
| Stability (30 min) | ±0.005 °C |
| Radial uniformity | ±0.01 °C at full range |
| Axial uniformity (60 mm) | ±0.035 °C at -40 °C; ±0.10 °C at 100 °C |
| Heating time | 15 min (23 to 160 °C) |
| Cooling time | 8 min (160 to 23 °C) |
| Immersion depth | 160 mm (6.30 in) |
| Insert OD | 31.9 mm (1.26 in) |
| Resolution | 0.001 °C |
| Display | 6.5 in color touch screen |
| Power | 90–254 VAC, 45–65 Hz, 580 W |
| Weight | 11.2 kg (24.7 lb) |
| Communication | USB A, USB B, RJ45, Wi-Fi, Bluetooth |
Manufacturer information: ADT878 datasheet on additel.com.