Ametek Jofra ETC400A115B11EX Temperature Calibrator, ETC-400 A Dry Block, 28 to 400 °C, 115 V AC
Ametek Jofra ETC400A115B11EX Temperature Calibrator
Overview
The ETC-400 A is built for speed: its heating block uses a highly efficient element with the insertion holes drilled directly into the block, keeping mass low for very fast heating and cooling. A bold checkmark stability indicator confirms when the set temperature is reached, with a 3-minute countdown timer alerting the operator before the stable condition.
The one-key-one-function interface gives immediate access to temperature setting and an auto-stepping function, so field technicians can run multi-point checks without menus. Output accuracy is +0.5 °C with the internal reference, stability is ±0.15 °C, and an external JOFRA STS reference probe can be used for tighter work. RS232 communication and JOFRACAL calibration software are part of standard delivery, and this model runs on 115 V AC.
Features
- 28 to 400 °C range at 23 °C ambient
- Heats up to 100 °C/min and stabilizes in about 3 minutes
- Accuracy +0.5 °C (internal reference); stability ±0.15 °C
- Stability indicator with 3-minute countdown timer
- One-key-one-function interface with auto-stepping
- RS232 interface with JOFRACAL calibration software included
- Fast cooling for rapid turnaround between test points
- Compact and light enough to carry in a tool box
Applications
- Field verification of temperature sensors and switches
- Calibration of thermocouples, RTDs and thermostats
- On-site maintenance of process temperature instrumentation
Ametek Jofra ETC400A115B11EX Temperature Calibrator Technical Specifications
| Temperature range | 28 to 400 °C (82 to 752 °F) at 23 °C ambient |
|---|---|
| Accuracy (internal reference) | +0.5 °C |
| Stability | ±0.15 °C (measured after 10 min of stability indication, 30 min measuring) |
| Display resolution | 0.1 ° or 1 °C/°F |
| Temperature units / slope rate | °C or °F; 0.1 to 9.9 °/min |
| Time to stability | approximately 3 minutes |
| Heating | 28 to 200 °C in 2 min; 200 to 400 °C in 3 min |
| Cooling | 400 to 200 °C in 9 min; 200 to 50 °C in 24 min |
| Reference probe | external JOFRA STS probe supported |
| Communications | RS232 with JOFRACAL software |
| Power | 115 V AC model (this configuration); other voltages available |
Manufacturer information: ETC Series on ametekcalibration.com.