Ametek PK2-854WC-SS/C Pneumatic Deadweight Tester, 4 to 854 inH₂O

Ametek PK2-854WC-SS/C Pneumatic Deadweight Tester, 4 to 854 inH₂O

$6,356.00
Skip to product information
Ametek PK2-854WC-SS/C Pneumatic Deadweight Tester, 4 to 854 inH₂O

Ametek PK2-854WC-SS/C Pneumatic Deadweight Tester, 4 to 854 inH₂O

$6,356.00

Ametek PK2-854WC-SS/C Pneumatic Deadweight Tester

Overview

The PK II is AMETEK's floating-ball pneumatic deadweight tester for low-pressure calibration. A ceramic ball and the weight stack float on a thin film of air inside the nozzle, so the generated pressure is essentially free of friction, and the flow regulator keeps the ball floating with no operator adjustment.

The Ametek PK2-854WC-SS/C Pneumatic Deadweight Tester pairs the standard ±0.05% accuracy grade with the /C suffix: its certificate includes effective area, nominal pressure and actual pressure data. The 4 and 10 inH₂O weight carriers and ball outputs provide fine steps across the whole water-column range.

Features

  • Floating-ball pneumatic design, virtually frictionless
  • 4 to 854 inH₂O certified range in 1 inH₂O steps
  • ±0.05% of reading accuracy
  • Certificate with effective area and pressure data (/C)
  • 4 and 10 inH₂O weight carriers and ball outputs
  • Rugged ceramic measuring ball
  • Self-regulating flow control, no valve adjustment
  • Closed-cover operation for outdoor use

Applications

  • Calibration of low-pressure gauges and transmitters
  • Draft and duct pressure instrument checks
  • Water-column instrument calibration

What's in the Box

  • Tester base with case
  • Stainless steel weight set
  • Tubing kit
  • Instruction manual
  • NIST-traceable certificate of calibration with data

Ametek PK2-854WC-SS/C Pneumatic Deadweight Tester Technical Specifications

Certified range 4 to 854 inH₂O
Increment 1 inH₂O
Weight carrier / ball output 4 and 10 inH₂O
Accuracy ±0.05% of reading
Certificate /C: effective area, nominal and actual pressure data
Repeatability ±0.005% of reading
Temperature coefficient ±0.00167% of reading per °C outside 23°C
Pressure source Nitrogen or instrument/shop air
Supply pressure 30 to 100 psi (206 to 689 kPa)
Connection 1/4 in female NPT with 1/4 in tube adapter
Ball Ceramic
Water column reference 20°C (ISA standard); 60°F AGA on request
Case Large case, 15 × 14.5 × 9 in
Weight Up to 28 lb (12.7 kg)

Manufacturer information: PK II on ametekcalibration.com.

You may also like