Fluke 1555 KIT Insulation Tester
Overview
Selectable test voltages run from 250 V to 1000 V in 50 V steps and above 1000 V in 100 V steps, covering the full range of test voltages in IEEE 43-2000. Dielectric Absorption Ratio and Polarization Index are calculated automatically.
A guard system cancels surface leakage current, live-circuit warning shows up to 600 V AC or DC on the display, and results are stored in up to 99 labeled memory locations. The 12 V lead-acid battery delivers more than 750 tests per charge and the instrument carries a 3-year warranty.
Features
- Test voltages 250 V to 10 kV
- Insulation resistance up to 2 T-ohm
- Automatic DAR and PI calculation
- Fluke Connect wireless remote operation and real-time trending
- Guard terminal eliminates surface leakage
- 99 memory locations with user-defined labels
- Leakage current 1 nA to 2 mA, capacitance 0.01 to 15 microF
- CAT III 1000 V / CAT IV 600 V, 3-year warranty
Applications
- Motor, generator and transformer insulation testing
- HV and MV cable, switchgear and bus testing
- Preventive maintenance programs with Fluke Connect reporting
What's in the Box
- Test cables with alligator clips (red, black, green)
- Heavy duty alligator clip leads
- AC power cord
- IP67 hard case
- Quick reference card
- USB-IR cable installation guide
- Certificate of calibration
Fluke 1555 KIT Insulation Tester Technical Specifications
| Test voltage (DC) | 250 V to 10,000 V |
|---|---|
| Insulation resistance range | Up to 2 T-ohm (10 kV: 200 G-ohm to 2 T-ohm range) |
| Accuracy (typical band) | 5% of reading on specified mid ranges; 20% on high ranges |
| Test voltage accuracy | -0%, +10% at 1 mA load |
| Leakage current | 1 nA to 2 mA, ±(5% + 2 nA) |
| Capacitance | 0.01 to 15.00 microF, ±(15% + 0.03 microF) |
| Timer | 0 to 99 minutes, 1 min setting resolution |
| Live circuit warning | 30 to 660 V AC/DC |
| Bar graph | 0 to 2 T-ohm |
| Display | Large digital/analog LCD, 75 x 105 mm |
| Power | 12 V lead-acid rechargeable battery, over 750 tests per charge |
| Safety | IEC 61010 CAT III 1000 V / CAT IV 600 V, IP40 |
| Operating temperature | -20 to 50 degC |
| Dimensions / weight | 269 x 277 x 160 mm, 3.7 kg |
Manufacturer information: Fluke 1555 Kit on fluke.com.