Skip to product information
Additel ADT878PC-425-NO-110V Dry Well + Process Calibrator 425C
$7,007.00
Additel ADT878PC-425-NO-110V Dry Well + Process Calibrator 425C
Overview
The ADT878PC-425 is an Additel 878 Reference Dry Well Calibrator in the 33 °C to 425 °C range, fitted with the Process Calibrator (PC) option. It combines a dual-zone temperature source with a two-channel readout for RTDs and thermocouples, a reference PRT input and a 6.5 in color touchscreen. The PC option adds voltage and current measurement, a 24 V output and a HART communicator function.
Features
- Dual-zone temperature control with a 6.5 in (165 mm) color touchscreen
- Setpoint control by internal or external reference sensor
- Self-calibration of the internal control sensor against the external reference
- Two RTD channels and one thermocouple channel, plus a 4-wire reference PRT input (ITS-90, CVD, IEC-751)
- Quick-Push connectors and 6-pin Lemo smart connector for the reference PRT
- HART communicator to read, configure and calibrate HART devices
- Documentation of up to 1,000 tasks with as-found and as-left results, plus data logging with auto step and ramp
- USB A, USB B, RJ45, Wi-Fi and Bluetooth communication
Applications
Stable heat source for calibrating temperature sensors such as RTDs and thermocouples, and for configuring and calibrating HART transmitters.
What's in the Box
- Reference dry well with selected insert (1)
- Power cable (1)
- USB cable (1)
- Insert removal tool (1)
- Thermal shield (1)
- Test leads, 2 sets (4 pcs)
- ISO 17025 accredited calibration certificate (1)
Additel ADT878PC-425-NO-110V Dry Well + Process Calibrator 425C Technical Specifications
- Temperature range: 33 °C to 425 °C (at 23 °C ambient)
- Display accuracy: ±0.2 °C full range
- Stability (30 min): ±0.010 °C at 100 °C; ±0.015 °C at 225 °C; ±0.020 °C at 425 °C
- Axial uniformity (60 mm): ±0.10 °C at 100 °C; ±0.15 °C at 225 °C; ±0.25 °C at 425 °C
- Radial uniformity: ±0.025 °C at 100 °C; ±0.030 °C at 225 °C; ±0.040 °C at 425 °C
- Hysteresis (display sensor): 0.04 °C
- Resolution: 0.001 °C; units °C, °F, K
- Heating time: 15 min (23 °C to 425 °C)
- Cooling time: 24 min (425 °C to 100 °C); 15 min (100 °C to 50 °C)
- Immersion depth: 193 mm (7.60 in); insert OD 30.8 mm (1.21 in)
- Reference PRT input: 0 to 50 Ω ±1.25 mΩ; 50 to 400 Ω ±0.0025% of reading; -200 °C to 962 °C
- RTD channels (2): 0 Ω to 4 kΩ; PT10 to PT1000, CU10, CU50, CU100, NI100, NI120
- Thermocouple types: S, R, K, B, N, E, J, T, C, D, G, L, U; -75 mV to 75 mV, accuracy 0.01% of reading + 5 µV
- Voltage measurement: ±12 V and ±30 V ranges, ±0.01% of reading + 0.6 mV
- Current measurement: -30 mA to 30 mA, ±0.01% of reading + 2 µA
- DC output: 24 V ±10%, maximum 60 mA
- Power: 90 to 254 VAC, 45 to 65 Hz, 1400 W (110 V or 220 V ordering option)
- Size (H x W x D): 170 x 345 x 330 mm (6.69 x 13.58 x 13.0 in)
- Operating environment: 0 °C to 50 °C, 0% to 90% RH non-condensing