Additel ADT878PC-425-NO-220V Temperature Calibrator, 33 to 425 °C Reference Dry Well with Process Calibrator Option

Additel ADT878PC-425-NO-220V Temperature Calibrator, 33 to 425 °C Reference Dry Well with Process Calibrator Option

$7,007.00
Skip to product information
Additel ADT878PC-425-NO-220V Temperature Calibrator, 33 to 425 °C Reference Dry Well with Process Calibrator Option

Additel ADT878PC-425-NO-220V Temperature Calibrator, 33 to 425 °C Reference Dry Well with Process Calibrator Option

$7,007.00

Additel ADT878PC-425-NO-220V Temperature Calibrator

Overview

The ADT878PC is the Process Calibrator version of Additel's 878 reference dry-well series, which covers -40 to 700 °C across three models. The PC option adds an external reference input plus a two-channel readout for RTDs and thermocouples, along with mA/VDC measurement and switch testing, turning the dry well into a complete temperature and process calibration station.

The 425 model covers 33 to 425 °C at 23 °C ambient (per the 2026 datasheet revision) with ±0.2 °C display accuracy across the full range and stability of ±0.010 °C at 100 °C. An automated self-calibration routine aligns the internal control sensor to an external reference, keeping long-term drift out of your uncertainty budget.

Features

  • Reference dry-well range 33 to 425 °C (at 23 °C ambient)
  • Display accuracy ±0.2 °C at full range
  • Stability ±0.010 °C at 100 °C; ±0.015 °C at 225 °C; ±0.020 °C at 425 °C (30 min)
  • Process Calibrator option: RTD and 2 TC channels, mA/VDC, switch test
  • Automated self-calibration of internal control sensor
  • Heats 23 to 425 °C in 15 minutes
  • External reference input for traceable comparison
  • 220 V AC power (model ADT878PC-425-NO-220V)

Applications

  • Reference calibration of RTDs and thermocouples
  • Calibration of process sensors and temperature switches
  • Laboratory temperature reference and comparison checks

Additel ADT878PC-425-NO-220V Temperature Calibrator Technical Specifications

Temperature range (at 23 °C ambient) 33 °C to 425 °C
Display accuracy ±0.2 °C at full range
Stability (30 min) ±0.010 °C at 100 °C; ±0.015 °C at 225 °C; ±0.020 °C at 425 °C
Radial uniformity ±0.025 °C at 100 °C; ±0.030 °C at 225 °C; ±0.040 °C at 425 °C
Axial uniformity ±0.10 °C at 100 °C; ±0.15 °C at 225 °C; ±0.25 °C at 425 °C
External reference sensor accuracy ±0.01 °C at 100 °C; ±0.02 °C at 425 °C
Heating time 15 min (23 to 425 °C)
Process calibrator inputs External RTD reference, 2 TC channels, mA/VDC, switch test
Insert / well diameter Insert OD 31.9 mm (1.26 in)
Environmental 8 to 38 °C guaranteed accuracy; 0 to 50 °C operating, 0-90% RH
Storage temperature -20 to 60 °C
Power 220 V AC (ADT878PC-425-NO-220V)

Manufacturer information: ADT878 datasheet on additel.com.

You may also like