Ametek MG PK2-654WC-SS-1B/C Deadweight Tester, 4 to 654 inH2O
Ametek MG PK2-654WC-SS-1B/C Deadweight Tester
Overview
The PK2-654WC-SS-1B/C is a Model PK II pneumatic deadweight tester that generates pressure by applying the force of its weight set over the effective area of a ceramic ball floating in a stainless steel cylinder. Because the ball floats on a thin film of air, operation is virtually frictionless and the weights never need rotating during a test.
A built-in flow regulator automatically adjusts the input air to hold the ball and weights in float, compensating for supply pressure variation and making the instrument accurate independent of the operator. The -1B accuracy option specifies ±0.025% of reading, and the /C option adds effective area, nominal pressure and actual pressure data for the delivered unit.
Features
- Range 4 to 654 inH2O with 1 inH2O increments; 4 and 10 inH2O carrier output
- ±0.025% of reading accuracy (-1B option)
- Calibration data included: effective area, nominal and actual pressure (/C)
- Virtually frictionless aluminum oxide ceramic measuring ball
- Self-regulating flow control, no operator adjustment between set points
- Stainless steel weights calibrated to standard or local gravity
- NIST traceable; overhung weight carrier reduces side thrust and friction
Applications
- Calibration of low-pressure gauges and transmitters in labs and field shops
- Natural gas industry instrumentation calibrated to ISA or AGA water columns
- Verification of pneumatic calibrators and pressure modules
What's in the Box
- Tester base with case
- Stainless steel weight set, range specific
- Instruction manual
- NIST traceable calibration certificate with data (/C)
Ametek MG PK2-654WC-SS-1B/C Deadweight Tester Technical Specifications
| Type | Pneumatic floating-ball deadweight tester |
|---|---|
| Range | 4 to 654 inH2O (approx. 0.14 to 23.6 psi) |
| Pressure increment | 1 inH2O |
| Weight carrier / ball output | 4 and 10 inH2O |
| Accuracy | ±0.025% of reading (-1B) |
| Repeatability | ±0.005% of reading |
| Temperature coefficient | ±0.0017% of reading/°C at 23 °C |
| Measuring ball | Aluminum oxide ceramic |
| Weight material | Stainless steel (SS); aluminum for small increments |
| Test connection | 1/4 in female NPT |
| Pressure source | Nitrogen or instrument/shop air (ISA quality) |
| Supply pressure | 30 to 100 psi (206 to 689 kPa) |
| Gravity | 9.80665 m/s² standard, or local gravity on request |
| Water column reference | 20 °C (ISA) standard; 60 °F (AGA) optional |
Manufacturer information: PK II on ametekcalibration.com.