Skip to product information
Ametek PK2-404WC-SS-1B/C Pneumatic Deadweight Tester, 4 to 404 inH₂O
$5,736.00
Ametek PK2-404WC-SS-1B/C Pneumatic Deadweight Tester
Overview
The PK II is AMETEK's floating-ball pneumatic deadweight tester for low-pressure calibration. A ceramic ball and the weight stack float on a thin film of air inside the nozzle, so the generated pressure is essentially free of friction, and the flow regulator holds the ball at float without operator input.
The Ametek PK2-404WC-SS-1B/C Pneumatic Deadweight Tester combines the -1B accuracy grade with the /C suffix: the certificate includes effective area, nominal pressure and actual pressure data. The 404WC is one of the two small-case PK II water-column models, with a 4 inH₂O weight carrier and ball output.
Features
- Floating-ball pneumatic design, virtually frictionless
- 4 to 404 inH₂O certified range in 1 inH₂O steps
- ±0.025% of reading accuracy (-1B grade)
- Certificate with effective area and pressure data (/C)
- Rugged ceramic measuring ball
- Self-regulating flow control, no valve adjustment
- Closed-cover operation for outdoor use
- Small case for bench or portable use
Applications
- Calibration of low-pressure gauges and transmitters
- Draft and duct pressure instrument checks
- Water-column instrument calibration
What's in the Box
- Tester base with case
- Stainless steel weight set
- Tubing kit
- Instruction manual
- NIST-traceable certificate of calibration with data
Ametek PK2-404WC-SS-1B/C Pneumatic Deadweight Tester Technical Specifications
| Certified range | 4 to 404 inH₂O |
|---|---|
| Increment | 1 inH₂O |
| Weight carrier / ball output | 4 inH₂O |
| Accuracy | ±0.025% of reading (-1B) |
| Certificate | /C: effective area, nominal and actual pressure data |
| Repeatability | ±0.005% of reading |
| Temperature coefficient | ±0.00167% of reading per °C outside 23°C |
| Pressure source | Nitrogen or instrument/shop air |
| Supply pressure | 30 to 100 psi (206 to 689 kPa) |
| Connection | 1/4 in female NPT with 1/4 in tube adapter |
| Ball | Ceramic |
| Water column reference | 20°C (ISA standard); 60°F AGA on request |
| Case | Small case, 15 × 10.5 × 7 in |
| Weight | Up to 28 lb (12.7 kg) |
Manufacturer information: PK II on ametekcalibration.com.