Ametek RK Pneumatic Deadweight Tester, 0.145 to 300 psi, ±0.05% of Reading

Ametek RK Pneumatic Deadweight Tester, 0.145 to 300 psi, ±0.05% of Reading

$5,251.00
Skip to product information
Ametek RK Pneumatic Deadweight Tester, 0.145 to 300 psi, ±0.05% of Reading

Ametek RK Pneumatic Deadweight Tester, 0.145 to 300 psi, ±0.05% of Reading

$5,251.00

Ametek RK Pneumatic Deadweight Tester

Overview

The AMETEK RK is a pneumatic deadweight tester: a primary pressure standard in which a ceramic ball floats on a film of clean air or nitrogen while stainless steel masses load it to generate pressure. Certified range models run from 1 to 51 psi (RK-50) up to 1 to 2011 kPa (RK-2000N), with pressure range 0.145 to 300 psi (1 to 2000 kPa) overall, so gauges and transmitters can be calibrated against a directly traceable force balance rather than a secondary reference.

Three accuracy grades are available: ±(0.05% of reading) as standard, ±(0.025% of reading) with the -1B suffix, and ±(0.015% of reading) with the -1A suffix, with repeatability ±0.005% of reading. Weights are stainless steel as standard, the ball is ceramic, the connection is 1/8 in female NPT, and standard delivery includes the tester base with case, weight set, manual and NIST-traceable certificate. Water-column models reference 20°C (ISA) or 60°F (AGA) and gravity is set to the international standard 9.80655 m/s² unless local gravity is specified.

Features

  • Accuracy options: ±0.05%, ±0.025% or ±0.015% of reading
  • Repeatability ±0.005% of reading
  • Pressure range 0.145 to 300 psi / 1 to 2000 kPa across models
  • Ceramic ball with stainless steel weight masses
  • Operates on nitrogen or instrument/shop air, 30 to 450 psi supply
  • Readings in psi, kg/cm², bar, kPa, inH2O, cmH2O and inHg
  • NIST-traceable certificate included as standard

Applications

  • Primary-standard calibration of pressure gauges and transmitters
  • Benchmark for digital pressure calibrators in metrology labs
  • Shop and field verification of low-pressure pneumatic instruments

Ametek RK Pneumatic Deadweight Tester Technical Specifications

Accuracy ±(0.05% of reading) standard; ±0.025% (-1B); ±0.015% (-1A)
Repeatability ±0.005% of reading
Temperature coefficient ±0.00167% of reading/°C outside 23°C
Pressure range 0.145 to 300 psi / 1 to 2000 kPa (family)
Gravity reference 9.80655 m/s² international standard; local gravity on request
Water column reference 20°C ISA standard; 60°F AGA standard when specified
Pressure source Nitrogen or instrument/shop air
Supply pressure 30 to 450 psi (210 to 3,100 kPa)
Output flow 1 scfh at 0.15 psi to 60 scfh at 300 psi
Connection 1/8 in female NPT
Ball / masses Ceramic ball; stainless steel masses (aluminum small increments optional)
Tester enclosure 15 × 9 × 8 in; 18 to 31.5 lb
Weights case 12 × 8.5 × 11.5 in; masses 17.5 to 61.5 lb
Standard delivery Tester base with case, weight set, manual, NIST-traceable certificate

Manufacturer information: RK deadweight tester datasheet on ametekcalibration.com.