Ashcroft AM2 Pressure Module

Overview

The AM2 family comes in two sensor types. AM2-1 low pressure modules use a capacitive sensor for differential and compound measurement from 0.25 to 200 in H2O on clean dry gas, with a maximum static line pressure of 100 psi. AM2-2 medium and high pressure modules use an isolated 316 stainless steel piezoresistive sensor for gauge, absolute, compound and vacuum ranges from 5 to 10,000 psi. Accuracy codes are ±0.025%, ±0.05% or ±0.1% of span (the 10,000 psi range is psig, ±0.1%), with overpressure margins of 200% up to 1,000 psi and 150% above that on AM2-2. Every module connects through a 1/8 in NPT female process connection, with an optional VCR fitting up to 5,000 psi.

Performance holds across the working week: repeatability is ±0.01% of span and sensitivity ±0.002% typical, and the standard temperature effect is ±0.004% of span per °F over the 20 to 120 °F (-7 to 49 °C) compensated range. Enhanced temperature compensation options keep ±0.025% accuracy with no additional ambient error. Each Quick-Select module ships with a traceable calibration certificate, and the interchangeable design lets you swap ranges on the ATE-2 without recalibrating the base unit.

Features

  • Quick-Select interchangeable modules for the ATE-2 handheld calibrator
  • AM2-1 low pressure: 0.25 to 200 in H2O, capacitive, differential and compound
  • AM2-2 medium/high: 5 to 10,000 psi, gauge, absolute, compound and vacuum
  • Accuracy options of ±0.025%, ±0.05% or ±0.1% of span
  • Repeatability ±0.01% of span; sensitivity ±0.002% typical
  • Compensated 20 to 120 °F; enhanced temperature compensation options
  • 316 SS isolated piezoresistive sensor; 1/8 in NPT female connection
  • Traceable calibration certificate included

Applications

  • Calibrating pressure transducers, transmitters and gauges
  • Pressure switch testing and pressure-decay leak checks
  • Field calibration where ranges change frequently

Ashcroft AM2 Pressure Module Technical Specifications

Family AM2 Quick-Select pressure modules (AM2-1 low, AM2-2 medium/high)
For use with Ashcroft ATE-2 handheld digital calibrator
AM2-1 ranges 0.25 to 200 in H2O, differential and compound
AM2-2 ranges 5 to 10,000 psi, gauge, absolute, compound and vacuum
Accuracy ±0.025%, ±0.05% or ±0.1% of span (range dependent)
Repeatability ±0.01% of span
Sensitivity ±0.002% typical
Temperature effect ±0.004% of span per °F (standard)
Compensated range 20 to 120 °F (-7 to 49 °C)
Overpressure (AM2-2) 200% to 1,000 psi; 150% above 1,000 psi
Process connection 1/8 in NPT female (optional VCR fitting to 5,000 psi)